Measurement object
Oxygen concentration in combustion exhaust gas and mixed gas
Measurement system Zirconium oxide sensor
Measurement range 0.01 to 100 vol% O2
Output signal 4 to 20 mA DC, two points (ZR402G) (maximum load resistance 550Ω)
Setting range
Any setting in the range of 0 - 5 vol%O2 to 0 - 100 vol%O2 (in 1 vol% O2), or partial range
Digital communication (HART)
250 to 550Ω, depending on number of field devices connected to the loop (multi-drop mode)
Sample gas pressure
-5 to +250 kPa (Pressure compensation option necessary over +5 kPa)
Sample gas temperature
0 to 700℃ (Probe only)
700 to 1871℃ (with 0.15m probe with JIS 5K 32 FF flange and ZR22P High Temperature Probe Adapter)
Probe length
0.15, 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0, 3.6, 4.2, 4.8, 5.4 m
Contact output
Four points (one is fail-safe, normally open)
Contact input
Two points
Calibration
Method; zero/span calibration
Calibration mode; automatic, semi-automatic and manual (with an LCD touchscreen)
Construction of detector (ZR22G)
Heater and thermocouple replaceable construction.
Non-explosion-proof. Equivalent to NEMA4X/IP66
Non-explosion-proof, outdoor installation, equivalent to NEMA4X/IP66
Ambient Temperature
Probe (ZR22G): -20 to 150℃
Reference Gas System
Instrument Air (preferred): 50 kPa (200 kPa with check valve) + the pressure inside the furnace; Approx. 1 Nl/min consumption
Natural convection
Power supply
85 to 264 V AC; 45 to 66 Hz
Performance of Yokogawa ZR22G-040-S-K-C-T-M-E-A ZR22G Zirconia Oxygen Analyzer Detector
Repeatability ±0.5% Maximum value of set range
Linearity ±1% Maximum value of set range
Drift ±2% Maximum value of set range/month
Response time (T90) 90% response within 5 sec (60 sec for explosion-proof ZR22S)