Manufacturer : Yokogawa
Product No. : ZR22G-200-S-C-E-T-T-E-A/CV
Product type : Separate type Zirconia Oxygen/ High Temperature Humidity Analyzer, Detector
Length : 200 = 2.0 m
Wetted material : S = Stainless steel
Flange : C = ANSI Class 150 4 RF SUS304 (JIS)
Reference gas : E = External connection (Instrument air)
Gas Thread : T = 1/4NPT(Female)
Connection box thread : T = 1/2 NPT
Instruction manual : E = English
A = Always -A
Options Valves : CV = Check valve (
Sample Gas Temperature: 0 to 700°C (Probe only) It is recommended to mount the cell using inconel cell-bolts when the temperature is greater than 600°C.
Sample Gas Pressure: - 5 to + 20 kPa (When the pressure in the furnace exceeds 3 kPa, it is recommended to use pressure compensated type. When the pressure in the furnace exceeds 5 kPa, pressure compensated type is required.) No pressure fl uctuation in the process should be allowed
Probe Length: 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0 m
Probe Material: JIS SUS316 stainless steel
Ambient Temperature: -20 to +150°C
Reference Gas System: Natural Convection, Instrument Air,or Pressure compensated
Wetted Material: SUS316 (JIS), Zirconia, SUS304 (JIS) (fl ange), Hastelloy B, (Inconel 600, 601)
Installation: Flange mounting
Weight: Approx. 17 kg
Shipping weight : 20 Kg